ICT / ITE / Medical Power Supply

GTM964500PWWWVV.V-F Image
Model ImageGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Information
Model NumberGTM964500PWWWVV.V-F
DescriptionGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Agency Documents
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O3a00000Ly9m5EAB
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000Ly9m5EAB
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal / carte/interne
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor / adaptateur pour alim à découpage régulée
CategoryICT / ITE / Medical Power Supply / ICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)6A
Max Output Power Rating (W)450
Output Voltage Range (V)12-54
Auxiliary/Fan Output Voltage (V)12
Auxiliary/Fan Output Power (W)15W Forced Air, 10W Convection
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionN/A
Size (mm)127.0 (L) x 76.2 (W) x 38.5 (H)
EnclosureUser-added image

Customer may need to provide EMI shielding if not in metal chassis
User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)VoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air)RFQ
* (Final ordering #/Part No to be defined)
Specifications

Features

  • High power density 3" x 5"� format
  • Auxiliary +12V fan output
  • Auxiliary +5V standby output
  • Power-good, remote-off, & remote-sense
  • Class I, Class II, or Class II with Functional Earth input configurations
  • ITE/ICT and medical (2 x MOPP) certified

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:5.8A typ. @ 90VAC, 2.3A typ. @ 230VAC (full load)
No Load Input Power:<150mW
Inrush Current:<35A @ @ 100VAC , <70A @ 230VAC (cold start @ 25°C)
Efficiency:85 - 95% typ. (full load)
Power Factor:>0.95 @ 115VAC, >0.90 @ 230VAC (full load)

Output

Output Power:For 12-18.9V output: 400W (25CFM forced air), 225W (convection), inclusive of fan/aux power
For 19-54.0V output: 450W (25CFM forced air), 250W (convection), inclusive of fan/aux power
Turn-on Delay:< 3 seconds @ 90VAC, < 2 seconds @ 115VAC
Output Voltage Tolerance:± 2.0% (worst case, set point + load/line regulation)
Load Regulation:± 0.5% (measured at connector)
Line Regulation:± 0.2% (measured at connector)
Ripple:1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:<5% (full load)
Hold-up Time:8ms typ. (full load)
+12V Fan Output:800mA (max)
+5V Standby Output:1A (max)

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-150%. Latched-off, cycle AC to reset
Over-Current Protection:105-150% typ. Non-latched with auto-recovery
Short Circuit Protection:Non-latched with auto-recovery
Over-Temperature Protection:Latched-off, cycle AC to reset

Environmental

MTBF:200,000 hours @ 25°C ambient
Operating Temperature:-20°C to 40°C (Full output power), -40°C to 80°C (Derated)
Storage Temperature:-40°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:NC: <100µA SFC: <500µA
Earth Leakage CurrentNC: <0.5mA SFC: <1.0mA
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact Discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 10V/m 80-2700MHz, 80% 1KHz AM.
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11

 

Derating CurveUser-added image
User-added image
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionUser-added image
Optional Input Wiring Kits are available:
PN/Ordering NoLengthEnd AEnd B
C810070M6VAR2(R)60IEC 60320 C8JST VAR-2
C1410070M6VAR2(R)60IEC 60320 C13JST VAR-2
VAR2210070M7VAR2(R)73JST VAR-2JST VAR-2
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
User-added imageUKCA Certification
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request

What is an open frame or internal AC-DC Power supply?

Open frame AC-DC power supplies are also known as AC-DC converters and are a type of internal power supply. The open frame AC to DC power supply is built into the inside of a system and used to convert the AC current into DC voltage.

Qu'est-ce qu'une alimentation IEC 62368-1: 2018?
La IEC 62368-1: 2018 s'applique à la sécurité des équipements électriques et électroniques dans le domaine de l'audio, de la vidéo, des technologies de l'information et de la communication et des machines professionnelles et de bureau avec une tension nominale ne dépassant pas 600 V. Il s'agit d'une norme de sécurité des produits qui classifie les sources d'énergie, prescrit des garanties contre ces sources d'énergie et fournit des orientations sur l'application et les exigences de ces garanties. Les mesures de protection prescrites visent à réduire la probabilité de douleur, de blessure et, en cas d'incendie, de dommages matériels. Un convertisseur ou un bloc d'alimentation AC / DC classé IEC 62368-1 est un bloc d'alimentation qui a été testé et certifié conforme à ladite norme.

Les blocs d'alimentation IEC 62368-1 vont-ils remplacé les blocs d'alimentation IEC 60950-1?
Le 20 décembre 2020, les normes de sécurité qui se sont historiquement appliquées aux équipements informatiques et audio-vidéo, IEC 60065 et IEC 60950-1, vont passées à la IEC 62368-1. En Amérique du Nord, les certifications ETL et UL 60950-1 resteront valides et pourront être utilisées jusqu'à nouvel ordre.

Quelle est la différence entre une alimentation IEC 62368-1 et une alimentation IEC 60950-1?
Bien que la structure et le contenu des normes soient très différents en général, la plupart des produits qui étaient auparavant conformes à la norme IEC 60950-1 sont conformes à la dernière norme IEC 62368-1 avec peu ou pas de modification.

Mon bloc d'alimentation IEC 60950-1 restera-t-il disponible?
GlobTek continuera à fabriquer et à fournir des alimentations électriques IEC 60950-1 indéfiniment jusqu'à nouvel ordre.

Qu'est-ce que l'efficacité de niveau VI?
Les chiffres I à VI sont un protocole international de marquage d'efficacité qui permet aux fabricants de désigner les performances d'efficacité d'un produit. Ces performances sont basées sur la mesure de l'efficacité énergétique de fonctionnement et de la consommation électrique à vide des alimentations externes ou des adaptateurs CA - CC testés à différents niveaux de charge.
La consommation électrique à vide fait référence à la quantité d'énergie consommée par une alimentation externe même lorsque l'appareil alimenté est éteint ou déconnecté.

L'efficacité de niveau VI est-elle requise?
L'efficacité de niveau VI est obligatoire dans certains pays et régions pour certains produits ou certaines applications.

Qu'est-ce que la directive éco-conception 2009/125 / CE?
La directive sur l'éco-conception a promulgué en Europe les dernières exigences en matière d'efficacité et de conservation énergétiques pour les adaptateurs CA - CC et les alimentations externes.

Comment fonctionne un adaptateur AC-DC ?

L'adaptateur CA / CC ou alimentation externe se compose d'une alimentation qui est enfermée dans un boîtier similaire à une prise secteur, d'où le terme de "plug type", "wall plug in", "plug-in" ou "plugtop power adapter". L'alimentation est branchée sur la prise secteur et la tension alternative est ensuite convertie par l'alimentation en une tension continue qui est délivrée à l'appareil via le câble de sortie.


Demande d'info

Model

Call
Email
💬
Text
💁
Chat
?
Demande de devis
Pouvons-nous vous aider?
Pouvons-nous vous aider?